CIM-3300P Permeameter for Feebly Magnetic Materials is designed according to《GB/T 35690-2017 Methods for the determination for of the relative magnetic permeability of feebly magnetic materials》and《ASTM A342/342M-14 Standard Test Methods for Permeability of Feebly Magnetic Materials. it can measure the relative permeability of feebly magnetic materials (for example austenitic stainless steel) with bar or square shape whose relative permeability µr is between 1.001-4.000. The system is controlled by computer, which can automatically complete the test and generate the test report.
Test Items
Range of Relative permeability µr : 1.001-4.000
Technical Data
1. Hardware Parameters
1.1 LZ-510 POWER SUPPLY
Model | LZ-510 |
Input Power | Single sheet 220V 50Hz |
Output Voltage | 0~±50V |
Output Current | ± 10A (Max value) |
Output protection | Over-voltage, over-current, short circuit protection. |
1.2 LZ-840 Fluxmeter
Model | LZ-840 |
Integrator | Range | 0.4, 4, 40, 400 mWb |
Measuring Accuracy(full range) | 0.25% |
Zero Drift | Less than 1μWb/60 secs |
Drift Adjust Method | Automatic, manual |
Drift Suppression | Intelligent suppression |
1.3 LXG-0630 Solenoid
Model | LXG-0530 |
Max. Current | 10A |
Max. Magnetic Field | 40000A/m@10A(higher magnetic field needs to be customized) |
Uniform Range | Diameter*Length Φ40mm*200, uniformity is superior to 0.5% |
1.4 PC-6616 Data Acquisition Card
Parameters | Specifications |
A/D convert | 16Bit, 8 channels differential input |
Convert time | < 10μs |
Convert accuracy | 0.01% (full range) |
D/A convert | 16Bit, 2 channels independent output |
Convert accuracy | 0.01%(full range) |
SMTest Software
SMTest runs in Windows 7 operating system, confirms to Windows software standards, easy to operate.
Chinese, English operating interfaces can be selected.
File system adopts data base, it can directly print or output test results into Excel.
File management with powerful functions, it can automatic save data, delete data and delete all data.
Data file contains full sampling data, sample parameters and instrument parameters. The file is in text format, and can input to other software.
Support all types of printers, test report and printer paper match perfectly.
With function of print preview, convenient to adjust the size of test report.
Directly print test report, or generate the test report into JPG file.
Send JPG file by E-mail, or save it into disk.
Test report contains full curves, test results, test conditions and sample parameters.
User logo and name can be added into test report.