MATS-313XSA Hysteresisgraph can automatically measure the soft magnets with low permeability, such as metal magnetic core , the measured parameters are: Ps, Bm, Hm, Ss, μa and δ, also μ′, μ″, μL, μR, Q and AL can be calculated under certain conditions.The test system is composed of industria...
MATS-313XSA Hysteresisgraph can automatically measure the soft magnets with low permeability, such as metal magnetic core , the measured parameters are: Ps, Bm, Hm, Ss, μa and δ, also μ′, μ″, μL, μR, Q and AL can be calculated under certain conditions.
The test system is composed of industrial computer, excitation power supply, high-precision wide-band power analyzer and power compensation module. The system adopts power compensation technology, and the maximum excitation power can reach to 3000VA or 6000VA, which effectively solves the problem that the power consumption of high-power and low-permeability soft magnetic devices cannot be measured.
Test Items
► Connect the measure coil to the voltage input port of standard power meter, calculate the peak value of magnetic flux density by reading the voltage effective value of the power meter.
Technical Data
Working Environment | |||||
Parameters | Specifications | ||||
Input Power | Single-phase 220V, 50Hz | ||||
Environment | Temperature: 23±5℃; Humidity: 30~75%RH | ||||
External magnetic field interference | Absolutely avoid | ||||
Thermal balance time | 10 minutes | ||||
TPS-500A Broadband Program Control Power Supply | |||||
Sub-model | 1207A | 2007A | 3007A | ||
Max. Output power | 500VA | 500VA (≤120k) | 500VA (≤120k) | ||
Frequency | 1kHz~120kHz | 1kHz~200kHz | 1kHz~300kHz | ||
Frequency fineness | < 0.1%×current value, frequency error< 0.05% | ||||
Output voltage | 0~10V~75V~150V~300V, four automatic ranges | ||||
Max. Output current | 7A | 7A | 7A | ||
1.√ means supports the function, Х means does not support the function 2.Model definition: max output power(VA), suffix means max frequency (kHz)/10+ max output current (A)+letter | |||||
LZ-300 Power Analyzer | |||||
Model | LZ-300/3010 | LZ-300/3020 | |||
Frequency | 1kHz~300kHz | ||||
Current | 0~10A (Effective value) | 0~20A (Effective value) | |||
Voltage | 0~300V (Effective value) | 0~120V(Effective value) | |||
Phase accuracy | 1% ±0.05° | 1% ±0.05° | |||
Model definition: max frequency kHz/10+max current A, add 0 in the front when the max current is less than 10. | |||||
ACR-100X Power Compensation Module | |||||
Model | ACR-100B | ACR-100 | |||
Frequency | 10k~300kHz | 10k~300kHz | |||
Max. Working voltage | 300V | 300V | |||
Max. Working current | 10A | 20A | |||
Max. Compensation power | 3000VA | 6000VA | |||
Compensation way | Automatic | Automatic | |||
Remarks: Power compensation is currently only used for Fixed Bm measurement and does not support fixed magnetic field measurement. |