BACK TO:About Linkjoin

CIM-3150MDT
CIM-3150MDT is a permanent magnets measurement device with high accuracy. It can fast measure flat or three dimensional magnetization angle of magnetized permanent magnets between magnetization vector and magnets fixed axis, meanwhile, it can measure the magnet moment of permanent magnets and has function of sorting. By measuring the magnetic flux signal of the magnet rotate core circle in the Helmholtz Coil to fast separate process and get the magnetic moment and angle.  
 
CIM-3150MDT has high accuracy and small angle error. It is different from the method of calculating magnetic declination angle by measuring the magnetic moment component of three axes. The drift, precision of integrator, verticality of three-dimensional Holmholtz coil and the accuracy of coil constant have great influence on the measurement of declination angle. The measurement accuracy of magnetic moment and the orthogonal precision of coil do not affect the measurement accuracy of the deflection angle, which can be applied to a wide range of samples of different sizes.
 





Test Items                                                 

 Test sample category: hard magnets such as NdFeB, SmCo, AlNiCo, Ferriet and so on.
Test sample: two poles magnetization.
Can directly measure the declination angle of hard magnets (need to equip with customized fixture).
Sample size: Determined by the uniform region of the Helmholtz coil.



 

Technical Data

Working Environment
Parameters Specifications
Input Power Single-phase 220V, 50Hz
Environment Temperature: 23±5℃; Humidity: 30~75%RH
External magnetic field interference Absolutely avoid
Thermal balance time 10 minutes
CIM-3150MDT  Permanent Magnets Declination Angle Measurement Device
Measure modes Explanations
One dimensional Angle The deflection angle of the horizontal component of the actual magnetization vector of the sample
Three dimensional Angle The deflection angles is obtained by α and β of the horizontal component of the actual magnetization vector.
LZ-841 Fluxmeter
Parameters Specifications
Header display 4¾ digits
Integrating capacitor 1 µF
Input resistance 100 k,10 k,10 k,10 k
Range 400mVs,40mVs,4 mVs,0.4mVs
Resolution 0.1,0.01s,0.001,0.0001 mVs
Accuracy Migration: ±4 µVs ±DC integrator drift;     
Gain:±0.25% reading(maximum change rate of input signal<10Vs/s)
Integrator drift Less than ±1µVs/minute



 

SMTest Test Software

► The software interface is mainly composed of: Menu bar, Graphic display bar, Data Bar and Parameters input and Control bar
► Export the measurement data displayed in the current measurement data bar in EXCEL format.
► Data table can display ID, product lot number, product material, product specification, 3D Angle, upper limit of deflection Angle, X component Angle, Y component Angle, magnetic moment, tester, date and remarks, etc.
► Support qualified judgment, can preset qualified range, the data marked red when the measurement results beyond the range.
► It can be searched by measuring time or recording parameters. Record parameters include those entered in "Product Lot Number", "Tester" and "Remarks".