MATS-313XSA Hysteresisgraph
MATS-313XSA Hysteresisgraph can automatically measure the soft magnets with low permeability, such as metal magnetic core , the measured parameters are: Ps, Bm, Hm, Ss, μa and δ, also μ′, μ″, μL, μR, Q and AL can be calculated under certain conditions.
The test system is composed of industrial computer, excitation power supply, high-precision wide-band power analyzer and power compensation module. The system adopts power compensation technology, and the maximum excitation power can reach to 3000VA or 6000VA, which effectively solves the problem that the power consumption of high-power and low-permeability soft magnetic devices cannot be measured.



 Test Items                                                 

1. Double winding measurement
► Shapes of test samples
Directly winding excitation (N1) and induction (N2) on samples to measure closed circuit samples with shape of ring, rectangle, EE, and EI.
► Category of test samples
Can test soft magnets such as Fe-Si-Al, Fe-Al, amorphous/nanocrystalline, silicon steel, permalloy, and soft ferrite.
► Test parameters
The measured parameters are: μa, δ, Ps, Br and Hc, and can calculate μ′, μ″, μL, μR, Q and AL under certain conditions.

2. Single winding measurement
The device supports single winding measurement and can directly measure magnetic devices such as single winding photovoltaic energy storage inductance and ring inductance. At this time, the power consumption Pe includes the line loss of the winding coil.


 

Features

► Adopt digital power supply and standard power meter which are controlled by computer and the whole measure process can be finished automatically.
► Digital programmable power supply has the function of automatically change the ranges, and can automatically match the output range and sampling range according to the material characteristics, which effectively improve the safety of equipment and measurement accuracy.
► Test samples with closed magnetic circuit and can be directly winded and measured on sample. Magnetizing coil (N1) and measuring coil (N2) to form a no-load transformer.
►Magnetizing coil circuit serial with the current of standard power meter, and get the magnetic field peak by reading the current peak value of power meter.
► Connect the measure coil to the voltage input port of standard power meter, calculate the peak value of magnetic flux density by reading the voltage effective value of the power meter.

Technical Data

Working Environment
Parameters Specifications
Input Power Single-phase 220V, 50Hz
Environment Temperature: 23±5℃; Humidity: 30~75%RH
External magnetic field interference Absolutely avoid
Thermal balance time 10 minutes
TPS-500A Broadband Program Control Power Supply
Sub-model 1207A 2007A 3007A
Max. Output power 500VA 500VA (≤120k) 500VA (≤120k)
Frequency 1kHz~120kHz 1kHz~200kHz 1kHz~300kHz
Frequency fineness < 0.1%×current value,  frequency error< 0.05%
Output voltage 0~10V~75V~150V~300V, four automatic ranges
Max. Output current 7A 7A 7A
1.√ means supports the function, Х means does not support the function
2.Model definition: max output power(VA), suffix means max frequency (kHz)/10+ max output current (A)+letter
LZ-300 Power Analyzer
Model LZ-300/3010 LZ-300/3020
Frequency 1kHz~300kHz
Current 0~10A (Effective value) 0~20A (Effective value)
Voltage 0~300V (Effective value) 0~120V(Effective value)
Phase accuracy 1% ±0.05° 1% ±0.05°
Model definition: max frequency kHz/10+max current A, add 0 in the front when the max current is less than 10.  
ACR-100X Power Compensation Module
Model ACR-100B ACR-100
Frequency 10k~300kHz 10k~300kHz
Max. Working voltage 300V 300V
Max. Working current 10A 20A
Max. Compensation power 3000VA 6000VA
Compensation way Automatic Automatic
Remarks: Power compensation is currently only used for Fixed Bm measurement and does not support fixed magnetic field measurement.
 
 

SMTest Test Software

► Running environment: Windows operating system.
► System language: Chinese or English.
► Multiply units can be selected to meet the use habits of different customers.
► Support input sample parameters with shapes of ring, EE, EI, CD, rectangle, diplopore, BS and others.
► Calculate Ae and Le according to sample sizes.
► Multiple groups of measurement conditions can be set at a time, such as fixed frequency, fixed Bm and fixed Hm can be set at the same time. Multi-condition test can be completed with one click.
► Can set 1 to 100 test points for each group at a time, quick setting in batches is supported.
► Save the text conditions to “Saved test conditions”, there is no need to duplicate input test conditions.
► The file system adopts database format and can output test results to Excel tables
► Test data is automatically saved, supporting historical data query, import and delete.