Sample, magnetizing coil (N1) and measuring coil (N2) to form a non-load transformer. It can measure the samples with closed magnetic circuit structure such as ring, rectangle, CD shape, EE shape, UU shape and PQ shape. | ||
Materials | Test Parameters | Applied Standards |
Silicon Steel, amorphous, nanocrystalline, permalloy and SMC Soft Magnetic Composite | Ps, Bm, Hm, Ss, ua, AC Magnetization Curve and Losses Curve |
GB/T 3658-2008 GB/T 19346.1-2017 |
Epstein Frame is a magnetizing device used to measure magnetic characteristics of electrical steel strip. It consists of four coils and air flux compensation coil, each of four coils contains a primary magnetization winding and a secondary induction winding. When testing, a primary coil, a secondary coil and a sample of electrical steel strip as an iron core to form a no-load transformer. | ||
Model | ES-700 | ES-200 |
Applied Standards |
GB3655-2008 and IEC 60404-2 |
GB 10129-88 |
Equivalent magnetic circuit | 940mm | 940mm |
Turns | N1=N2=700 turns | N1=N2=200 turns |
Applied frequency | 50~400Hz | 400Hz~10kHz |
Measured range |
Oriented silicon steel sheet: Bm1.0T~1.8T,Hm≤1000A/m Non-oriented silicon steel sheet: Bm0.8T~1.5T,Hm≤10kA/m |
|
Sample size | Width: 30mm±0.2mm; length: 280~320mm±0.5mm | |
Sample mass | 240g~1000g | 200g〜1000g |
The Permeameter is a magnetizing device used in the measurement of a single piece of electrical steel strip. It is composed of yoke, exciting coil and pole shoe. During the test, the tested sample is put into the Permeameter and with the magnetic yoke to form a closed magnetic circuit. The Permeameter can be customized according to customer requirements. |
|
Measured parameters | Specifications |
Applied Standards | GB13789-2008 and IEC 60404-3 |
Applied frequency range | 50Hz~150Hz |
Measured range |
Oriented silicon steel sheet: Bm1.0T~1.8T, Hm≤1000A/m Non-oriented silicon steel sheet: Bm0.8T~1.5T, Hm≤10kA/m |
Sample size | 500mm X 500mm (Single sheet) |
Sample mass | 240g~1000g |
Working Environment | ||||
Parameters | Specifications | |||
Input power | Single-phase 220V, 50Hz | |||
Environment | Temperature: 23±5℃; Humidity: 30~75%RH | |||
External magnetic field interference | Absolutely avoid | |||
Thermal balance time | 10 minutes | |||
MATS-3100M Hysteresisgraph | ||||
Model | 0505A | 1205A | 2005A | 1020A |
Max. Output power |
500VA(20〜40Hz:150VA; 40〜1KHz: 500VA; 1K〜2K Hz:350VA) |
2000VA(≤400Hz) | ||
Frequency | 40Hz~500Hz | 40Hz~1200Hz | 40Hz~2000Hz | 45Hz~1kHz |
Frequency fineness | 1Hz | 1Hz | 1Hz | 1Hz |
Output voltage | 0~1V~10V~50V~150V four ranges | 0~10V~50V~80V~100V four ranges | ||
Max. Output current | 10A | 10A | 10A | 20A |
Sampling rate | 2MHz | 2MHz | 2MHz | 2MHz |
Resolution and linearity | 16Bit | 16Bit | 16Bit | 16Bit |
Standard interfaces | RJ45, RS-232 | |||
Current method | √ | √ | √ | √ |
Waveform compensation | √ | √ | √ | √ |
Harmonic analysis | Х | Х | √ | Х |
1.√ means support this function, Х means do not have the function.. 2. Model definition: max frequency (Hz)/100+max power(VA)/100+letter(A: with current method; B:with current method and H coil method) |
System Technical Index | ||||||||
► According to National Standard GB/T 3655-2008, under the frequency of 50Hz, 60Hz, test 30*300mm silicon steel reference sample by 25cm Epstein Frame: | ||||||||
Measured parameters | Ps(%) | Hrms(%) | Bm(%) | Hm(%) | ||||
Uncertainty (k=2) | 1 | 2 | 1 | 2 | ||||
Repeatability (contact temperature) | ± 0.5 | ± 1 | ± 0.5 | ± 1 | ||||
► According to National Standard GB/T 13789-2008, under the frequency of 50Hz, 60Hz, test 500*500mm silicon steel reference sample by SST-500 Permeameter: | ||||||||
Measured parameters | Ps(%) | Hrms(%) | Bm(%) | Hm(%) | ||||
Uncertainty (k=2) | 1.5 | 2 | 1 | 3 | ||||
Repeatability (contact temperature) |
± 0.5 | ± 1 | ± 0.5 | ± 1 | ||||
► According to National Standard GB/T 3658-2008, under the frequency of 20Hz~2kHz, test permalloy ring sample: | ||||||||
Measured parameters | Ps(%) | μa(%) | δ(%) | Bm(%) | Hm(%) | |||
Uncertainty (k=2) | 3 | 2 | —— | 1 | 1 | |||
Repeatability (contact temperature) |
± 1.5 | ± 1 | ± 1 | ± 0.5 | ± 0.5 | |||
► According to National Standard GB/T 19346.1-2017, measure amorphous ring sample: | ||||||||
Measured parameters | Ps(%) | μa(%) | Ss(%) | Bm(%) | Hm(%) | |||
Uncertainty (k=2) | 3 | 3 | 3 | 1 | 1 | |||
Repeatability (contact temperature) |
± 1.5 | ± 1.5 | ± 1.5 | ± 0.5 | ± 0.5 |